연구성과
알림마당
Development of an immunoaffinity chromatography and LC-MSMS method for the determination of 6 zearal 조회수 : 472 2019-05-13 |
A novel and simple method for detecting 6 zearalenones in animal feed using liquid chromatography coupled with triple quadrupole mass spectrometry (LC-MS/MS) and immunoaffinity columns (IAC) was developed. The chromatographic peaks of the 6 zearalenones were successfully identified by comparing their retention times and mass spectrum with reference standards. The mobile phase was composed of mobile phase A (water) and B (0.5% formic acid in ACN). Method validation was performed with linearity, sensitivity, selectivity, accuracy and precision. The limits of detection (LODs) for the instrument used to study zearalenones ranged from 0.3 to 1.1 μg/kg, and the limits of quantification (LOQs) ranged from 1.0 to 2.2 μg/kg. Average recoveries of the 6 zearalenones ranged from 82.5% to 106.4%. Method replication resulted in intra-day and inter-day peak area variation of <3.8%. The developed method was specific and reliable and is suited for the routine analysis of zearalenones in animal feed. |
5. 논문_제랄레논 이성질체.pdf |
- 이전글 Analytical method development and monitoring of Aflatoxin B1, B2, G1, G2 and Ochratoxin A in animal
- 다음글 Simultaneous determination of synthetic food additives in kimchi by liquid chromatography-electrospr